Glucose · L1 Assayed chemistry control lot CC-4471 (exp Mar 2027) · Erba EM 200 · target 92 ± 2.8 mg/dL

+3SD 100.4+2SD 97.6+1SD 94.8x̄ 92−1SD 89.2−2SD 86.4−3SD 83.608 Sep 07:30 · 92.8 mg/dL · z 0.308 Sep09 Sep 07:30 · 90.3 mg/dL · z -0.610 Sep 07:30 · 94.5 mg/dL · z 0.911 Sep 07:30 · 92.3 mg/dL · z 0.111 Sep12 Sep 07:30 · 88.6 mg/dL · z -1.213 Sep 07:30 · 93.1 mg/dL · z 0.414 Sep 07:30 · 98.4 mg/dL · z 2.3 · 1-2s14 Sep15 Sep 07:30 · 91.4 mg/dL · z -0.216 Sep 07:30 · 94 mg/dL · z 0.717 Sep 07:30 · 89.5 mg/dL · z -0.917 Sep18 Sep 07:30 · 93.4 mg/dL · z 0.519 Sep 07:30 · 95.1 mg/dL · z 1.120 Sep 07:30 · 90.9 mg/dL · z -0.420 Sep21 Sep 07:30 · 92.6 mg/dL · z 0.222 Sep 07:30 · 98.7 mg/dL · z 2.4 · 1-2s23 Sep 07:30 · 99.3 mg/dL · z 2.6 · 2-2s23 Sep24 Sep 07:30 · 92.8 mg/dL · z 0.325 Sep 07:30 · 90 mg/dL · z -0.726 Sep 07:30 · 93.7 mg/dL · z 0.626 Sep27 Sep 07:30 · 91.2 mg/dL · z -0.328 Sep 07:30 · 94.2 mg/dL · z 0.829 Sep 07:30 · 89.2 mg/dL · z -129 Sep30 Sep 07:30 · 92.3 mg/dL · z 0.101 Oct 07:30 · 93.1 mg/dL · z 0.4
Date / timeValuezWestgardByCorrective action · acknowledged by
08 Sep 07:3092.80.3in controlS. Kavitha
09 Sep 07:3090.3-0.6in controlS. Kavitha
10 Sep 07:3094.50.9in controlS. Kavitha
11 Sep 07:3092.30.1in controlS. Kavitha
12 Sep 07:3088.6-1.2in controlS. Kavitha
13 Sep 07:3093.10.4in controlS. Kavitha
14 Sep 07:3098.42.31-2sS. Kavitha
15 Sep 07:3091.4-0.2in controlS. Kavitha
16 Sep 07:30940.7in controlS. Kavitha
17 Sep 07:3089.5-0.9in controlS. Kavitha
18 Sep 07:3093.40.5in controlS. Kavitha
19 Sep 07:3095.11.1in controlS. Kavitha
20 Sep 07:3090.9-0.4in controlS. Kavitha
21 Sep 07:3092.60.2in controlS. Kavitha
22 Sep 07:3098.72.41-2sS. Kavitha
23 Sep 07:3099.32.62-2sS. KavithaRecalibrated glucose, fresh control vial, both levels rerun in range; 6 patient samples since last good QC rerun, no reportable change. · Dr. Priya Nair
24 Sep 07:3092.80.3in controlS. Kavitha
25 Sep 07:3090-0.7in controlS. Kavitha
26 Sep 07:3093.70.6in controlS. Kavitha
27 Sep 07:3091.2-0.3in controlS. Kavitha
28 Sep 07:3094.20.8in controlS. Kavitha
29 Sep 07:3089.2-1in controlS. Kavitha
30 Sep 07:3092.30.1in controlS. Kavitha
01 Oct 07:3093.10.4in controlS. Kavitha